¤ýTest target: Support up to 62" of sample size for LED BLU, CCFL BLU, LCD module, OLED module, PDP, etc
¤ýTest items : Color gamut, flicker rate, response rate (Optional) including luminance,chromatility, color temperature, Peak wavelength, dominant wavelength, Full width half max, Color rendering index, etc
¤ýMeasuring accuracy: wavelength ¡¾3%, chromatility ¡¾0.002
¤ýMeasuring method: Takes 4 minutes for 42" 1000 points including 0.1 second for movement and 0.1 seconds for measurement
¤ýMeasuring method : Measure using spectrophotometer after recognizing location via semi-automatic sample loading and vision
¤ýFeatures : For BLU consisting of RGB LED, it analyzes all optical properties of each LED of RGB and white mixed with three colors by only one measurement (Patent)
- Measure accurate and highly reliable luminance using integrating sphere type input optical system and optical power meter and color correction
- High speed and low noise by using clean room type linear motor
- Optimal for evaluating LED BLU and LED display
¤ýHigh resolution spectrophotometer(TE cooled 2048 CCD Sensor)
¤ýMeasure highly reliable optical properties by using 2" integrating sphere type light receiving head, optical power meter, and spectrometer at the same time
¤ýMeasure accurate and repeatable luminance by correcting color in real time using spectrometer
¤ýStandard measurement solid angle and distance via zoom lense
High speed and low noise linear motor and X-Y-Z robot
¤ýHigh speed and low noise by using clean room type linear motor
¤ýAccurately find and move to measuring location using Vision system to minimize tolerance
¤ýMeasure height of module using height sensor and by moving light receiving head to standard condition location
¤ýLoad various sizes of samples using multi-purpose sample jig
¤ýSemi-auto sample feeding
¤ýBlock light by maintaining measurement dark room
¤ýSW to provide various ranges of information.
¤ýUser is able to select and edit measuring items.
¤ýIntuitive user environment to easily recognize measurement process and result.
(Spectrum, color coordinate, current location, map, etc)
¤ýSeparated display for measurement of R,G,B,W
¤ýVarious motion mode: recognize measurement location using vision.
¤ýRecognize measurement location using user's coordinate file.
¤ýRecognize measurement location by user settings. (4, 5, 9, and 16 points, etc)
¤ýMeasurement date is save in informative form (CSV format): Total by each item and spectrum by each sample.
¤ýRGB item in sample: Report generation and save in accordance with user template.
¤ýDisplay overall distribution, set grade, process statistics, and calculate uniformity through map and data screen.
¤ýCIE and KS standard measurement mode and value
¤ýCorrection by using Korea Research Institute of Standards and Science certified correction lamp and LED
¤ýHigh speed and precision Keithley source meter and 2048 CCD sensor
(Select TE Cooled type) : high speed and resolution
¤ýProxy for Korea Research Institute of Standards and Science(KRISS) correction certificate (optional)
¤ýTechnical support and maintenance
- Customer oriented set up, user instruction and development of SW as per customer requirements
- Provide measurement knowledge and information through activities in CEI and photometry club
Service and prompt A/S
¡¡SPECIFICATION | |
Spectrometer |
400 to 900nm, TE cooled 2048 CCD array type |
Photodiode |
380 to 950nm, 10§± Area |
Robot System |
X-Y-Z Robot, Linear Motor, Resolution 1§ |
PD Temperature |
30¡É |
Max Sample size |
52 inch |
Measurement Speed |
0.1 sec / point |
Movement Speed |
0.1 sec / point |
Vision system |
CCD camera |
Dimension |
3,950mm(W) x 1,660mm(D) x 1,827mm(H) |